Basic specifications (Accuracy guaranteed for 1 year) |
|
Measurement object |
Covered electric conductors (minimum 100 mm2 ), Metallic terminals
Note: cannot measure shielded electric wires |
AC Voltage 40 to 66 Hz |
30.0V to 420.0V, 380V to 600V, 2ranges
(Basic accuracy: ±1.5% rdg. ±5 dgt. /true RMS rectified, band width 40Hz to 400Hz at 420.0V range) |
Crest factor |
1.8 (only sine wave at 600V range) |
Max. rated voltage to earth |
600V AC rms |
Functions |
Data hold function, auto-power save, low battery alarm |
Display |
Digital LCD, max. 4200 digit, (zero suppress at 10 digits or less) |
Sampling rate |
1.67 times/sec (approx. 600 msec period), Response time 2.4 sec or less |
Power supply |
AA alkaline (LR6) battery ×6, Continuous use: 14 hr (Power ON, no measurement), or 1-minute measurements for 1000 times, or 5-minute measurements for 200 times |
Dimensions, mass |
51mm (2.01in) W × 275mm (10.83in) H × 37.5mm (1.48in) D (one probe), 670g (23.6oz) |
Supplied accessories |
Instruction manual ×1, AA alkaline battery (LR6) ×6, Carrying soft case ×1 |
|
|
Basic accuracy Covered (unshielded) electric wires (Indoor PVC or equivalent, 100 mm2 or larger) |
|
Measurement range |
Accuracy guaranteed range |
Accuracy |
40 - 66 Hz |
66 - 400 Hz |
420 V |
30.0 - 420.0 V |
±1.5 % rdg. ±5 dgt. |
±2.5 % rdg. ±5 dgt. |
600 V |
380 - 480 V |
±2.0 % rdg. ±5 dgt. |
|
481 - 600 V |
±5.0 % rdg. ±5 dgt. |
|
|
Note: For insulated cables over 38 mm2 to less than 100 mm2, add 0.5% rdg. to measurement accuracy |
|